Micro Controlled based X-ray Timer
Features
- Microprocessor based x-ray timer.
- Independent memory for digital and conventional dental radiography time selections.
- Short time exposure range starting from .02 sec. makes it ideal for digital radiography.
- Soft touch panel switches.
- Line voltage compensation 170 - 260 V in 12 uniform steps.
- Tube overload protection for longer tube life.
- 10 seconds time delay between two consecutive exposures.
- Audio-visual indications during exposure and in tube overload condition.
- LED indicators to indicate ready sign and digital radiography (rvg) selection.
- Lead coated, irremovable, long radiography cone.
- Side mounting of tube head provides high maneuverability and quick positioning.
Technical Specifications
Tube voltage
|
70 kVp
|
Tube current
|
8 mA
|
Focal spot
|
0.8 mm
|
Beam Limiting device
|
Length 20cm, Diameter 6cm
|
Equivalent filtration
|
2.0mm of AI
|
Leakage radiation
|
<28 mR/Hr. at 1m distance
|
Timer
|
0.02 - 0.6 seconds for digital radiography 0.1 - 3.0 seconds for conventional radiography
|
Power supply
|
Single phase 230V, 50Hz
|
Maximum power
|
1500 W
|
- To
adapt to every installation condition, Ardent 70 can be configured with
30cm, 60cm, or 80cm extension arm giving a sweep range upto 205cm.
- The smooth movement of scissor arm makes positioning simple and easy.
- Ardent 70 is available in several configurations including wall mounts, mobile units and ceiling mounts.
|