X-Ray System: Medilux
300 generator is designed to fulfill the requirements of all
Radiography, Fluoroscope, Spot Filming and Special procedures with
manual or motorized table. The generator is suitable for use with image
intensifier TV systems.
Controls:
- Line switches: ON & OFF push button switches control line contractor.
- Voltage
compensator switch: A 15 position switch for voltage compensation in
the range 180 to 250volts. Voltmeter indicator at red bond indicates
compensated voltage at 230 volts.
- Major KVp Control: A 12 position switch for coarse selection of KVp.
- Minor
KVp Control: A 12 position selector switch for fine selection of KVp
enable selection of 1 KVp per step in range of 40 to 100 KVp.
- Fluoroscopy
KVp Selector: A 12 position independent switch permits selection of
fluoroscope KVp values from 45 to 100 KVp in step of 5KVp.
- mA
technique control: Rotary switch permits one of the following mA
selections 50mA, 100mA on small focus and 200mA, 300mA on large focus.
- Tube selector: A toggle switch with 'F' & 'R' option selects under couch or over couch X-Ray tube.
- Radiographic Timer: A 24-position switch permits exposure time selections from 0.02 to 10 seconds.
- Fluoroscopic
Timer: It permits operator to limit fluoroscopic examination to any
integrated time upto 5 minutes. Approx. 20 seconds before termination
of preselected time: timer will initiate a buzzer.
- Fluaro mA Selector: Continuously variable fluoro mA knob permits selection of mA from 0-3 mA.
- Fluoroxcopy exposure switch: Fluoro test.
Indicators:
- Digital read out for KVp & mAS Radiography/Spot Film exposure parameters.
- Selected Facal Spot is indicated on display penal.
- X-Ray LED glows during exposure period.
- MA
meter: Radiographic actual mA value is indicated on 0 - 350 mA analogue
scale and actual fluoro mA value is indicated on 0-10 mA scale.
- Tube overload LED glows in case selected factors are in excess of X-Ray tube exposure rating.
Power Line Requirement: Power
demand at rated output: Single phase, 50HZ, and 30 KVA 3 phase to
single pahse transformer is recommended, maximum allowable line
resistance in 415 line is 0.5 Ohms.
Compatible with: Motorized or Monual Table, Spot Film Device and IITV.
|