<< Previous   300 / 500mA, 125kVp - Diagnostic X-ray Generator

Product Code : 12

Product Specification
300 / 500mA, 125kVp - Diagnostic X-ray Generator
Medilux 325/525 - Generator is designed to fulfill the requirement of all Radiography, Fluoroscopy, Spot Filming and special procedures with manual or motorized tables. The generator is suitable for use with Image Intensifier TV systems.

Ergonomic Design Features
  • Single Phase (2 Pulse) X-Ray Generator, Maximum Rating 300/500 mA at 125 kVp (24/40KW) as per IS 7620:1986
  • Sleek control console with Digital Display of mA, mAs & kVp
  • High Voltage Silicon Rectifiers for consistent results and low system losses.
  • Instantaneous Tube Overload Protection
  • Spot filming possible on all mA stations
  • Compatible with BEL DRA-1/25/20/40 X-Ray Tube
Controls
  • Line Switches: ON & OFF push button switches
  • Voltage Compensator Switch: A 15 way switch compensate voltage in the range 180 to 250 volts.
  • Major kVp Control: A 12 position switch for coarse selection of kVp. Approx. 10 kVp per step.
  • Minor kVp Control: A12 position switch enables selection of 1 kVp per step in range of 45 to 125 kVp.
  • Fluoroscopy kVp Selector: A12 position independent switch permits selection of fluoroscopy KVp Values from 45 to 100 kVp in steps of 5 kVp.
  • Technique Control: 50, 100 & 150 mA on Small Focus and 200 mA, 300 mA on Large Focus in MEDILUX-325; 100, & 200 mA on Small Focus & 300 mA, 400 mA, & 500 mA on Large Focus in MEDILUX-525.
  • Tube Selector: A Toggle switch with 'F' & 'R' option selects undercouch or overcouch X-Ray tube.
  • Radiographic Timer: A 24 position switch permits exposure time selections from 0.01 to 5 seconds.
  • Fluoroscopic Timer: It limits fluoroscopic examination to an integrated time up to 5 minutes.
  • Fluoro mA Selector: Continuously variable fluoro mA knob permits selections of mA value from 0-3 mA.
  • Fluoroscopy Exposure Switch: Fluoro Test.
Indicators
  • Digital display of mA, mAS & kVp
    Radiography / Spot Film exposure parameters.
  • Selected Focal Spot, X-Ray and Tube overload LED's Display Penal.
  • mA meter: Radiographic actual mA value is indicated on 0-500 mA analogue scale and actual fluoro mA value is indicated on 0-10 mA scale.
Power Line Requirement
Power Demand at rated output Single Phase, 50 HZ, and 30/50 KVA. 3 Phase to single phase transformer is recommended Maximum allowable line resistance in 415 line is 0.5/0.2 Ohms.

Compatible With: Motorized or Manual Table, Spot Film Device and IITV
Other Products in 'Diagnostic X-Ray Systems ' category
RF X-ray Machine With Motorised Table
RF X-ray Machine With Motorised Table
Horizontal Bucky Table compatible with 60/500 mA
Horizontal Bucky Table compatible with 60/500 mA
Control Panel
Control Panel
60mA,100KVp Light Mobile X-Ray Machine
60mA,100KVp Light Mobile X-Ray Machine
X-Ray With Fluoroscopy Tilting Table
X-Ray With Fluoroscopy Tilting Table
Spot Film Device with 300/500 mA Units
Spot Film Device with 300/500 mA Units
 
View all products(12)
 
<< Back




 
Contact Details
 
MEDILUX SYSTEMS
E-1, 11/1, Sector-15, Rohini,Delhi - 110085, India
Phone:91-11-27883915/27883163
Fax:91-11-27883163
Send Mail   Send SMS   
Key Personnel
Mr. R. Singla (Proprietor)
Mobile:+919810257790, +919215537443







Copyright © 2011-12 by MEDILUX SYSTEMS All Rights Reserved.